Other, other text: d10.Device available for evaluation; h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.One device was received in "medium" condition with "wear and tear".Visual inspection noted the plug from the power cord was damaged and black meaning that a short circuit happened.The fuse from the power cord was also missing.The printed circuit board (pcb) and the fuses from the pcb were "ok".As the power cord was very damaged functional testing could not be performed in its current state.A new power cord was installed, and the device was fully functional.The temperature was stable and in range.A safety test was also performed, and all the values were acceptable.The root cause was due to a "blown" line cord fuse from suspected surge or shorted outlet.It was unknown what caused the problem.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced the power cord, tank cover and gasket and the o-rings.Device passed all functional and delivery tests.
|