Model Number HL-90
Device Problems Grounding Malfunction (1271); Electrical Shorting (2926)
Patient Problem Insufficient Information (4580)
Event Date 02/15/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the device had blown a fuse during surgery.Patient involvement is unknown.
 
Manufacturer Narrative
Other, other text: d10.Device available for evaluation; h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.One device was received in "medium" condition with "wear and tear".Visual inspection noted the plug from the power cord was damaged and black meaning that a short circuit happened.The fuse from the power cord was also missing.The printed circuit board (pcb) and the fuses from the pcb were "ok".As the power cord was very damaged functional testing could not be performed in its current state.A new power cord was installed, and the device was fully functional.The temperature was stable and in range.A safety test was also performed, and all the values were acceptable.The root cause was due to a "blown" line cord fuse from suspected surge or shorted outlet.It was unknown what caused the problem.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced the power cord, tank cover and gasket and the o-rings.Device passed all functional and delivery tests.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17192272
MDR Text Key317762583
Report Number3012307300-2023-06595
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public50695085820909
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/23/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberHL-90
Device Catalogue NumberHL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/11/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-