• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 PUREWICK FEMALE EXTERNAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 PUREWICK FEMALE EXTERNAL CATHETER Back to Search Results
Model Number PWFX30
Device Problem Increase in Suction (1604)
Patient Problem Pain (1994)
Event Date 04/14/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event Description
It was reported that the patient was experiencing pain with suction from using purewick female external catheter.Representative advised to stop usage and speak with doctor and the lot information not given.No medical intervention was reported.
 
Manufacturer Narrative
The reported event is inconclusive as no sample was returned for evaluation.It was unknown whether the device had met specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.A potential root cause could be due to "dimensions not specified correctly".The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "setup: 1.Connect the canister to wall suction and set to a minimum of 40mmhg continuous suction.Always use the minimum amount of suction necessary.If using the drydoctm vacuum station, connect the canister to the unit and turn the unit on.Please consult the drydoctm vacuum station user guide for further information.2.Connect the canister to wall suction and set to a minimum of 40mmhg continuous suction.Always use the minimum amount of suction necessary.If using the purewick¿ urine collection system, connect the canister to the unit and turn the unit on.Please consult the purewick¿ urine collection system user guide for further information.3.Using standard suction tubing, connect the purewick¿ female external catheter to the collection canister" h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event Description
It was reported that the patient was experiencing pain with suction from using purewick female external catheter.Representative advised to stop usage and speak with doctor and the lot information not given.No medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUREWICK FEMALE EXTERNAL CATHETER
Type of Device
PUREWICK FEMALE EXTERNAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key17198965
MDR Text Key317931046
Report Number1018233-2023-04657
Device Sequence Number1
Product Code NZU
UDI-Device Identifier00801741189050
UDI-Public(01)00801741189050
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberPWFX30
Device Catalogue NumberPWFX30
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/12/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
-
-