Model Number 8011-0501-01
Device Problem Failure to Deliver Shock/Stimulation (1133)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation evaluated the internal handles and the customer's report was not replicated or confirmed.The internal handles was put through extensive testing using known good spoons and connected to a test r series, without duplicating the report.The handles passed the continuity testing from the three-month check-out procedure in the internal handles operator's guide.The internal handles were scrapped as a precaution.A replacement set was sent to the customer.No trend is associated with reports of this type.
 
Event Description
Complainant alleged that during biomed testing, the associated device failed to discharge using these attached internal handles.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOCLAVABLE INTERNAL HANDLES (MSERIES)
Type of Device
INTERNAL PADDLE
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key17200295
MDR Text Key317853611
Report Number1220908-2023-02374
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K963781
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 06/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number8011-0501-01
Device Catalogue Number8011050101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2023
Was the Report Sent to FDA? No
Date Manufacturer Received06/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-