Model Number 1458Q/86
Device Problems Fracture (1260); Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/05/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented during implant procedure.During procedure, it was noted that the left ventricular lead has fractured and could not be separated from the guidewire.The reported lead was not installed and the procedure was completed with an alternative lead on (b)(6) 2023.There were no patient consequences.
 
Manufacturer Narrative
The reported events of lead fracture and failure to separate guidewire were confirmed.As received, a complete lead with stuck stylet was returned in one piece for analysis.A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.The ptfe coating of the stuck stylet was found bunched up with the inner coil.The connector pin was found pulled out consistent with procedure damage.The cause of the reported events was isolated to the bunching of the stylet ptfe coating inside the inner coil.The reported event of fracture was not found on the lead but excessive forces resulted in the connector pin to be pulled out and become separated from the connector.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17206139
MDR Text Key317987761
Report Number2017865-2023-24309
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeBR
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberA000119777
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/11/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GUIDEWIRE
-
-