• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PROGENIKA BIOPHARMA, S.A ID CORE XT; ANTIGEN GENOTYPING TEST

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PROGENIKA BIOPHARMA, S.A ID CORE XT; ANTIGEN GENOTYPING TEST Back to Search Results
Catalog Number 1020220034
Device Problem False Negative Result (1225)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/28/2023
Event Type  malfunction  
Manufacturer Narrative
The genomic dna sample was sent to grifols laboratory solution for dna sequencing.Next generation sequencing interrogated specific sequences: rh proximal promoter, exons 1-10, and portions of intron 2-3.The rh genotype detected by sequencing is a hybrid sequence of rhd exon2-intron2-exon3 within rhce gene.This hybrid sequence explains the failure of id core xt to detect the 109bp insert in intron 2 resulting in a false negative prediction of c antigen in discrepancy with serology result.The effect of this novel hybrid allele on rhce antigen expression is unknown and further review of serology data is recommended.This false negative result obtained by id core xt is considered a discrepant result and then a malfunction.This case report is associated with limitations of id core xt assay described in the id core xt package insert (limitations 1 and 9.7).
 
Event Description
It was reported that the sample (b)(6), from "centro de transfusión de la comunidad valenciana", was tested with serology.The test result was positive (c+), which contrasted with the molecular typing performed on (b)(6) 2023, using the id core xt assay which provided negative results (c-) with id core xt analysis software v3.0.2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ID CORE XT
Type of Device
ANTIGEN GENOTYPING TEST
Manufacturer (Section D)
PROGENIKA BIOPHARMA, S.A
ibaizabal bidea, edificio 504
parque tecnologico de bizkaia
derio, bizkaia 48160
SP  48160
Manufacturer Contact
jokin amo
ibaizabal bidea, edificio 504
parque tecnologico de bizkaia
derio, bizkaia 48160
SP   48160
MDR Report Key17217759
MDR Text Key318125115
Report Number3006413195-2023-00015
Device Sequence Number1
Product Code PEP
UDI-Device Identifier08437013457064
UDI-Public(01)08437013457064(17)240207(10)0203000029
Combination Product (y/n)N
Reporter Country CodeSP
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/07/2024
Device Catalogue Number1020220034
Device Lot Number0203000029
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/29/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/11/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-