• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OAKDALE LEVEL 1 FAST FLOW FLUID WARMER; WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 FAST FLOW FLUID WARMER; WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION Back to Search Results
Model Number 8002950
Device Problems Overheating of Device (1437); Temperature Problem (3022); Key or Button Unresponsive/not Working (4063)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/12/2023
Event Type  malfunction  
Event Description
It was reported that upon assembly and performance testing it was found that the power off button was not functioning.The device would over temp during testing.No patient involvement was reported.
 
Manufacturer Narrative
Other, date of event is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
Additional information received 5-july-2023 via email: issue was discovered prior to the patient use and no patient or clinical injury.There is no medical intervention required and the outcome of event is ongoing.The event occurred on 12- june-2023.
 
Manufacturer Narrative
Evaluation codes: updated device evaluation: one device was returned for investigation.Device failed flow rate test.Visual inspection found the device was missing the tank fill plug, had a cracked return tube, and the power off button does not work.The low flow rate can cause the device to "over temp" over time.It did not "over temp" during the investigation.The customer complaint was confirmed.Root cause was attributed to the power off button not working due to a broken membrane switch."over temp" caused by low flow rate due to the cracked return tube.What caused the damage could not be established.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.No actions taken at this time.Investigation only and device will be repaired when needed.Device will be sent to loaner pool.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 FAST FLOW FLUID WARMER
Type of Device
WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17221755
MDR Text Key318070621
Report Number3012307300-2023-06833
Device Sequence Number1
Product Code BSB
UDI-Device Identifier50695085829506
UDI-Public50695085829506
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK020043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 10/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8002950
Device Catalogue Number8002950
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/13/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-