Catalog Number 2C4041 |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/07/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that a y-type tur/bladder irrigation set was detected with particulate matter inside the sterile fluid path.This was discovered prior to patient use.There was no patient involvement.
|
|
Manufacturer Narrative
|
E1: initial reporter last name: (b)(6).E1: initial reporter address: (b)(6).G1: device manufacturer address 1: 600 mts.Oeste de entrada, principal ave.Las americas, parque industrial.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.A visual inspection of the photograph was performed which observed a particulate of the tip protector inside the tubing.The reported condition was verified.The cause of the reported condition was a manufacturing issue; potentially related to a misalignment at the automatic assembly fixtap per an incorrect location of the components.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|