Catalog Number 10825468001 |
Device Problems
High Test Results (2457); Calibration Problem (2890); Non Reproducible Results (4029)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/01/2023 |
Event Type
malfunction
|
Event Description
|
There was an allegation of questionable gen.2 ise indirect for na results for 2 patient samples on a cobas 6000 c (501) module.The initial results were reported outside of the laboratory.The results were questioned and the samples were repeated.For patient 1, the initial na result was 168 mmol/l.The repeat result was 142 mmol/l.For patient 2, the initial na result was 162 mmol/l.The repeat result was 132 mmol/l.The repeat results were deemed correct.
|
|
Manufacturer Narrative
|
The customer's qc was acceptable.The customer performed calibration after the questionable results were produced and received a calibration error.The investigation is ongoing.H3 other text : na.
|
|
Manufacturer Narrative
|
The field service engineer (fse) found a solenoid valve was malfunctioning and was causing insufficient cleaning of the cuvette cells.The fse also found a hole in the vacuum tubing, the fse replaced the valve and the tubing, ran ise checks, and performed calibration and qc successfully.The service maintenance actions performed by the fse resolved the issue.No further issues were reported after the service visit.
|
|
Search Alerts/Recalls
|