H4: device manufacture date: the lot was manufactured from january 19, 2023, to january 25, 2023.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.A visual inspection of the photograph provided noted that the tubing line was cut in two pieces.When the photo images were zoomed in, the cut mark appeared to be jagged and not smooth, which suggested the tubing line may have been stuck between two objects; then the user tried to pull the stuck tubing line away from the two objects and therefore resulted in the tubing line being broken in two pieces.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|