SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ANESTHESIA BREATHING SINGLE LIMB CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C1038J |
Device Problems
Crack (1135); Fracture (1260); Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/01/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the connection with the filter was cracked and leaked.No adverse patient effects were reported by the customer.
|
|
Manufacturer Narrative
|
B3: month and year of event have been provided, day is unknown.D4: lot number, expiration date, udi number and h4: device manufacture date is unknown, no information has been provided to date.G5: 510k is blank, device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
Device evaluation: one device was returned for investigation.Visual inspection confirmed the broken connector defect, however the failure did not cause leakage.Root cause was determined to be the broken connector.This issue could be generated by a coupe in the component.The most probable cause of this condition was improper handling of the product outside of manufacturing facilities.No lot number was provided therefor a device history report (dhr) review could not be completed.No corrective actions taken at this time.
|
|