Model Number 471205-17 |
Device Problems
Thermal Decomposition of Device (1071); Melted (1385)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Based on the claim against the product by the customer noting a thermal damage, an investigation is in progress to determine the cause of the reported event.A return material authorization (rma) was issued to the customer requesting to have the intuitive device returned.However, isi has not received the product involved with the alleged issue to perform failure analysis.As such, the probable root cause cannot yet be determined based on the information provided.Additional information is being gathered to determine the contribution of the device to the customer reported issue.A follow-up mdr will be submitted if the product is returned (post failure analysis evaluation) or if additional information is received.
|
|
Event Description
|
It was reported that during a da vinci-assisted myomectomy surgical procedure, the plastic holder of the fenestrated bipolar forceps instrument was melted.The procedure was completed with no reported injury.Intuitive surgical, inc.(isi) followed up with the initial reporter and obtained the following additional information: the fenestrated bipolar forceps instrument was inspected before use and there was no damage or anything unusual.At the instrument jaws where the cables run along the yellow plastic holder, this area was burned.The insulation of the conductor wire was not damaged.The instrument main tube was not damaged.The customer could not state exactly if the instrument tips did collide into other instruments or tools during the procedure.There was no cautery loss associated with the defect.There was no allegation of arcing, and the procedure was performed with a spare instrument.
|
|
Manufacturer Narrative
|
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The fenestrated bipolar forceps instrument was analyzed and reported failure was confirmed.The instrument was found to have the thermal damage on the bipolar yaw pulley.The instrument passed an electrical continuity test.The complaint regarding thermal damage was confirmed by failure analysis, which indicates that the device did contribute to the customer reported issue.
|
|
Event Description
|
Refer to h10/h11 for follow-up information.
|
|