• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC ENDOWRIST; FENESTRATED BIPOLAR FORCEPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC ENDOWRIST; FENESTRATED BIPOLAR FORCEPS Back to Search Results
Model Number 471205-17
Device Problems Thermal Decomposition of Device (1071); Melted (1385)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/01/2023
Event Type  malfunction  
Manufacturer Narrative
Based on the claim against the product by the customer noting a thermal damage, an investigation is in progress to determine the cause of the reported event.A return material authorization (rma) was issued to the customer requesting to have the intuitive device returned.However, isi has not received the product involved with the alleged issue to perform failure analysis.As such, the probable root cause cannot yet be determined based on the information provided.Additional information is being gathered to determine the contribution of the device to the customer reported issue.A follow-up mdr will be submitted if the product is returned (post failure analysis evaluation) or if additional information is received.
 
Event Description
It was reported that during a da vinci-assisted myomectomy surgical procedure, the plastic holder of the fenestrated bipolar forceps instrument was melted.The procedure was completed with no reported injury.Intuitive surgical, inc.(isi) followed up with the initial reporter and obtained the following additional information: the fenestrated bipolar forceps instrument was inspected before use and there was no damage or anything unusual.At the instrument jaws where the cables run along the yellow plastic holder, this area was burned.The insulation of the conductor wire was not damaged.The instrument main tube was not damaged.The customer could not state exactly if the instrument tips did collide into other instruments or tools during the procedure.There was no cautery loss associated with the defect.There was no allegation of arcing, and the procedure was performed with a spare instrument.
 
Manufacturer Narrative
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The fenestrated bipolar forceps instrument was analyzed and reported failure was confirmed.The instrument was found to have the thermal damage on the bipolar yaw pulley.The instrument passed an electrical continuity test.The complaint regarding thermal damage was confirmed by failure analysis, which indicates that the device did contribute to the customer reported issue.
 
Event Description
Refer to h10/h11 for follow-up information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST
Type of Device
FENESTRATED BIPOLAR FORCEPS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key17238757
MDR Text Key318200875
Report Number2955842-2023-16722
Device Sequence Number1
Product Code NAY
UDI-Device Identifier00886874119808
UDI-Public(01)00886874119808(10)K10230212
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K214095
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial,Followup
Report Date 06/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number471205-17
Device Catalogue Number471205
Device Lot NumberK10230212 0099
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/09/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
Patient SexFemale
-
-