DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
|
Back to Search Results |
|
Model Number N/A |
Device Problems
Thermal Decomposition of Device (1071); Failure to Power Up (1476)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/06/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is complete, a follow up/final report will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported that outside of surgery the unit did not power on at all due to a burnt power inlet module.There was no patient involved and no impact to patient and/ or user.Due diligence information complete.No additional information available.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under: (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record determined no power due to burnt power inlet module.Power inlet module was replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.Upon review of our reporting decision on blown fuses exhibiting signs of overheating.A review of all complaints for this product indicates that there has never been an event that has resulted in a death, serious injury, and/or actual fire when the fuse has blown or fuse drawer deformation.Therefore, zimmer biomet will not be reporting this failure mode going forward.
|
|
Event Description
|
No additional event information available.
|
|