TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 500; MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500
|
Back to Search Results |
|
Model Number 500AHCT |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/28/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during use of the device for cardiopulmonary bypass (cpb), the blood parameter monitor (bpm) venous values were inaccurate after calibration.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was a 45¿50-minute delay.There was no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
Updated block: b5.
|
|
Event Description
|
Per clinical review: on (b)(6) 2023, the team experienced a problem with their heart lung machine (hlm) while on cardiopulmonary bypass (cpb), described as 'venous not reading correctly even after calibration'.There was no blood loss, the unit was not changed out, and the procedure was completed successfully.The complaint was originally described as having a change-out with a 40¿50-minute delay, but after confirming with the user facility, this was not the case as they were referring to how long it took to borrow a unit from an affiliate hospital, and the unit was not changed out during that specific procedure.
|
|
Manufacturer Narrative
|
The reported complaint was not verifiable since the product was not returned for evaluation or testing.Multiple diligence attempts for part return were unsuccessful.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|