ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; NO MATCH Back to Search Results
Model Number 2088TC/52
Device Problems Failure to Capture (1081); Difficult to Insert (1316); Low impedance (2285); Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/14/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented for dual chamber pacemaker implant procedure.The novel atrial lead was reported to exhibit difficulty in setting up for implant.Upon positioning, the lead was noted to exhibit a pacing issue, low pacing impedance, and p-wave amplitude variation.The procedure was completed without implanting an atrial lead on (b)(6) 2023.There were no patient consequences.
 
Manufacturer Narrative
The reported events of low pacing impedance, p-wave amplitude variation, difficulty in setting up for implant were not confirmed as received, a complete lead was returned in one piece.Final analysis did not find any indication of conductor fractures or internal shorts.No anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17240257
MDR Text Key318303863
Report Number2017865-2023-24851
Device Sequence Number1
Product Code NVN
UDI-Device Identifier05414734502887
UDI-Public05414734502887
Combination Product (y/n)Y
Reporter Country CodeTH
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2088TC/52
Device Catalogue Number2088TC-52
Device Lot NumberP000164983
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/13/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-