Model Number 2088TC/52 |
Device Problems
Failure to Capture (1081); Difficult to Insert (1316); Low impedance (2285); Device Sensing Problem (2917)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/14/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the patient presented for dual chamber pacemaker implant procedure.The novel atrial lead was reported to exhibit difficulty in setting up for implant.Upon positioning, the lead was noted to exhibit a pacing issue, low pacing impedance, and p-wave amplitude variation.The procedure was completed without implanting an atrial lead on (b)(6) 2023.There were no patient consequences.
|
|
Manufacturer Narrative
|
The reported events of low pacing impedance, p-wave amplitude variation, difficulty in setting up for implant were not confirmed as received, a complete lead was returned in one piece.Final analysis did not find any indication of conductor fractures or internal shorts.No anomalies were found.
|
|