Device evaluation summary: the reported inconsistent channel issue was verified during service.The service technician reviewed the error codes in the instrument memory and noted nothing abnormal.The codes were all user related codes ie: (10-bad reference at start of test, 12-inactive channel and 15-au test aborted).The issue was resolved by cleaning and calibrating the lift wire assy to manufacture specifications.Conclusion: complaint is confirmed for the act plus instruments inconsistent channel issue.The issue was verified during service when the technician reviewed the error codes in the instrument memory and noted nothing abnormal.The issue was resolved by cleaning and calibrating the lift wire assy to manufacture specifications.Preventive maintenance / testing was performed per specifications.No patient/clinical safety issues reported.Trends for issues with this product are reviewed at quarterly quality meetings.This regulatory report is being submitted as part of a retrospective review and remediation per d00953163 as part of a capa action.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
Medtronic received information that during use this act plus instrument passed wet quality control, however, when testing the patient sample the instrument would stop at 25 on one channel and 33 on the other channel several times in a row.The instrument was changed out with a backup and there was no resulting adverse patient effect.
|