PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC EFFICIA DFM100; DEFIBRILLATOR Back to Search Results
Model Number 866199
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/06/2023
Event Type  malfunction  
Event Description
This report is based on information provided by philips field service engineer (fse) and has been investigated by the philips complaint handling team.Philips received a complaint on the efficia dfm100 defibrillator indicating that the no waves were visible, only dots.The fse evaluated the device on site.It was determined that this was a malfunction of the 5 lead ecg trunk and 3 leadset, grabber.The fse replaced these parts in order to resolve the reported issue.The device remains at the customer site and no further evaluation is warranted at this time.Based on the information available and the testing conducted, the cause of the reported problem was a malfunction of the 5 lead ecg trunk and 3 leadset, grabber.The reported problem was confirmed.Based on the information available and results of additional analysis, no further action is necessary at this time.The data entered in this complaint record will be utilized for product quality and safety improvements per the post market surveillance and risk management processes.The fse replaced the 5 lead ecg trunk and 3 leadset, grabber to resolve the reported issue.It has been concluded that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EFFICIA DFM100
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC
no.2 keji north 3rd road
nanshan district
shenzhen
CH 
Manufacturer (Section G)
PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC
no.2 keji north 3rd road
nanshan district
shenzhen
CH  
Manufacturer Contact
jacqueline nishino
no.2 keji north 3rd road
nanshan district
shenzhen 
CH  
7552698099
MDR Report Key17252232
MDR Text Key318506193
Report Number3030677-2023-02632
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K110825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/03/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866199
Device Catalogue Number866199
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/06/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-