• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PERFUSION SYSTEMS BIO CONSOLE 560 USER INTERFACE; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PERFUSION SYSTEMS BIO CONSOLE 560 USER INTERFACE; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number A09075001
Device Problems Display or Visual Feedback Problem (1184); Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/24/2021
Event Type  malfunction  
Manufacturer Narrative
Device evaluation summary: the reported instrument turned off and it's screen went blank were not verified during service.The service technician was unable to duplicate the reported issues.The service technician found error code 68 in the error log and this is indicative of an issue with the communication between the user interface and the base.As a precaution the technician replaced the bitsy xb pcb.Preventive maintenance was performed per specifications.Conclusion: complaint not confirmed for the reported user interface turning off and screen going blank during service.There were no patient/clinical safety issues reported.The service history record was not reviewed as returned product analysis found no evidence of servicing issues with the serviced device.Trends for issues with this product are reviewed at quarterly quality meetings.This regulatory report is being submitted as part of a retrospective review and remediation per (b)(4) as part of a capa action.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that during use this bio-console user interface (ui) instrument turned off during the case and it's screen went blank.The instrument was replaced with a backup and there was no adverse patient effect reported.Additional information received confirmed hand crank was not required to maintain flow at any stage during the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIO CONSOLE 560 USER INTERFACE
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key17258026
MDR Text Key318683656
Report Number2184009-2023-00784
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080824
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 07/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/05/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA09075001
Device Catalogue NumberA09075001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/24/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-