The device was returned and evaluated, and the customer¿s allegation (peeling of the tissue pad) was confirmed.Device evaluation found the tissue pad was worn and partially peeled.A review of the device history record confirmed the device had no abnormalities related to the reported event.It has been less than one year since the subject device was manufactured.Based on the results of the investigation, the root cause of the partially peeled off tissue pad could not be determined.A likely cause of the event could be output of ultrasonic waves with the grasping part closed (including after the tissue was cut).Not grasping the tissue could cause the tissue pad to wear and partially peel.The issue is addressed in the instructions for use (ifu): do not activate output while the grasping section is closed without contacting tissue or vessel or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.When cutting or vessel sealing is performed, apply light tension on the tissue so that users can confirm that they are transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.Olympus will continue to monitor the field performance of this device.
|
The customer reported to olympus that when using a sonicbeat 5mm, 35cm, front-actuated grip, part of the tissue pad peeled off without any shedding.The event occurred during an unspecified procedure and was completed with a similar device.There was no patient harm associated with the event.
|