• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF UNIT "ESG-400"; ELECTROSURGICAL GENERATOR ESG-400

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF UNIT "ESG-400"; ELECTROSURGICAL GENERATOR ESG-400 Back to Search Results
Model Number WB91051W
Device Problems Defective Component (2292); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/07/2023
Event Type  malfunction  
Manufacturer Narrative
The device evaluation was unable to confirm the customer¿s reported problem of loosing energy.However, an e433 error occurred due to a defective high voltage power supply board.The evaluation also noted a cracked front panel at the bottom right corner.The recorded fault log found multiple errors of e433, e006, e486 and e187.No further information regarding the event was provided by the customer.The dhr showed that the device was manufactured according to valid instructions and met all specifications.The investigation is still in progress; however, if additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The customer reported that during procedure, the high frequency electrosurgical generator was ¿losing energy¿.The device was returned for service and during the evaluation, the following reportable malfunction was identified: an e433 error due to a defective hvps board.The registered nurse at the user facility further reported the therapeutic robotic assisted bio-prostatectomy was completed with the same device without delay.The staff that performed the procedure stated the error code flashed on and off quickly and were unable to capture the exact code.No death or injury and no impact to patient or other has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF UNIT "ESG-400"
Type of Device
ELECTROSURGICAL GENERATOR ESG-400
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, de 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17264044
MDR Text Key318451959
Report Number9610773-2023-01805
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K203682
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 07/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWB91051W
Device Catalogue NumberWB91051W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/20/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-