Catalog Number 110003451
Device Problems Degraded (1153); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/12/2023
Event Type  malfunction  
Event Description
It was reported that during a procedure, the threaded inserter shaft was damaged and unable to hold the trial, impactor head and then was stuck unto the inserter.There as no harm or health consequences to the patient.It was reported that no further information is available.
 
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2023-01493.Canada.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified the devices were assembled upon return.A screwdriver was able to turn the threaded shaft such that it was able to disengage from the modular inserter.Discoloration was observed on the shaft of the modular inserter and around the etching.The strike plate exhibits impact marks.Scratching is also present on the shaft.The complaint is confirmed based on the evaluation of the returned product.Dhr was reviewed and no discrepancies related to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 STR MODULAR SHELL INSERTER
Type of Device
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17265873
MDR Text Key318516422
Report Number0001825034-2023-01492
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 08/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number110003451
Device Lot Number539710
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Date Manufacturer Received08/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/13/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-