PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD; INTERVENTIONAL FLOUROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number ALLURA XPER FD20
Device Problem Unstable (1667)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/08/2023
Event Type  malfunction  
Event Description
It has been reported to philips that the table pivot will not lock into place.The device was inside of clinical use at the time of the event.It was reported that the patient fell off the table due to the issue, but that no harm was sustained due to the fall.Philips has started an investigation of this complaint.
 
Manufacturer Narrative
Philips has investigated this complaint.A philips field service engineer (fse) inspected the system onsite and found that the patient table pivot movement is defective.The brakes appear to be no longer locking when they should be locking.Fse disassembled the table covers and removed pivot brake assembly.Fse installed new pivot brake assembly and reassembled table covers.Pivot brake assembly is now engaging and functional.After the replacement, the system was returned to use in good working order.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD
Type of Device
INTERVENTIONAL FLOUROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key17267633
MDR Text Key318558639
Report Number3003768277-2023-03957
Device Sequence Number1
Product Code OWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130842
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD20
Device Catalogue Number722012
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/22/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-