Model Number PE074F5
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/26/2023
Event Type  malfunction  
Manufacturer Narrative
The reported event of balloon issue was confirmed.As received, the catheter tip was bent.Balloon was found to be ruptured around circumference.Distal side of the balloon latex inverted to catheter tip.After pulling the inverted balloon latex back to the original position and cutting the proximal side of the balloon latex, edges of latex did not appear to match up.No other visible damage was found from the catheter body and windings.Further evaluation regarding related quality issues is under investigation.The lot number was not provided thus a device history record was not reviewed.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Event Description
It was reported that the balloon failed to maintain its inflation during balloon test before use.The issue was resolved by replacing the catheter.There were no patient complications reported.Corrected information was obtained from the condition of the returned product and the sales rep.It was confirmed that the event occurred on the first day of use since blood was observed inside the balloon.The brand and size of the used introducer is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION
Type of Device
ELECTRODE, PACEMAKER, TEMPORARY
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
jonathan diaz
1 edwards way
irvine, CA 92614
MDR Report Key17273028
MDR Text Key318587842
Report Number2015691-2023-14377
Device Sequence Number1
Product Code LDF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K813521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberPE074F5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/20/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/30/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-