The initial reporter stated they received questionable results for samples from one patient tested with the elecsys anti-ccp assay on a cobas e 801 analytical unit (serial number (b)(6)) and compared to the siemens method.An interference is suspected by the customer.On (b)(6)2023, a sample from the patient was tested using the elecsys anti-ccp assay, resulting in a value of 309 u/ml.On (b)(6) 2023, a sample from the patient was tested using the elecsys anti-ccp assay, resulting in a value of 249 u/ml.On (b)(6) 2023, a sample from the patient was tested using the siemens anti-ccp method, resulting in a value of <0.5 ku/l.This result is consistent with the patient¿s clinical picture.
|