Model Number 71020
Device Problem Malposition of Device (2616)
Patient Problem Inadequate Pain Relief (2388)
Event Date 06/13/2023
Event Type  Injury  
Event Description
Patient was implanted with the nalu peripheral nerve stimulator system on (b)(6) 2022.After activation the patient reported persistent communication issues between the implantable pulse generator (ipg) and the external therapy disc that were unable to resolved through conservative troubleshooting with nalu tech support and onsite representatives.On (b)(6) 2023 the physician performed a surgical procedure to relocate the ipg to a more superficial location in order to improve communication with the therapy disc.No components were explanted or replaced during the procedure and the patient reports no further communication issues.
 
Manufacturer Narrative
There are no allegations of system or component failure.All components remain in use with good results for the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NALU NEUROSTIMULATION SYSTEM
Type of Device
PERIPHERAL NERVE STIMULATOR
Manufacturer (Section D)
NALU MEDICAL INC
2320 faraday avenue
suite 100
calsbad CA 92008
Manufacturer (Section G)
NALU MEDICAL, INC
2320 faraday ave
suite 100
carlsbad
Manufacturer Contact
terry villarba
2320 faraday avenue
suite 100
carlsbad, CA 92008-7241
7604482360
MDR Report Key17282901
MDR Text Key318644046
Report Number3015425075-2023-00103
Device Sequence Number1
Product Code GZF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K183579
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/07/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71020
Device Catalogue Number71020
Device Lot NumberA04190
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/28/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-