• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) Back to Search Results
Model Number CP120I/200/004
Device Problems Leak/Splash (1354); Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/13/2023
Event Type  malfunction  
Manufacturer Narrative
Other, other text: d4: udi, expiration date, g5: no 510k as item is not sold in us, h4 - no information received to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that leaks were found during the pre-test.
 
Manufacturer Narrative
Device evaluation: one device and five (5) photos were returned for investigation.Upon visual inspection, a tear was found in the breathing bag confirming the customer complaint.The device was functionally tested where it was found that the hole did not affect the function of the bag.The customer found this issue prior to use.Further inspection showed that the hole was created by molding during the suppliers manufacturing process.Production personnel have been informed of this issue and will continue to monitor this issue.Device history review of the reported lot number shows no non-conformities during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX ANESTHESIA BREATHING CIRCUIT
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17289543
MDR Text Key318843858
Report Number3012307300-2023-07044
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCP120I/200/004
Device Catalogue NumberCP120I/200/004JP
Device Lot Number230123
Was Device Available for Evaluation? No
Date Returned to Manufacturer06/20/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/31/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-