Model Number CLV-S190
Device Problem Flare or Flash (2942)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/06/2023
Event Type  malfunction  
Manufacturer Narrative
E1/establishment name: organization for occupational health and safety (b)(6)hospital (added here due to character limitation).The device was returned to olympus for evaluation, and the customer's allegation was confirmed.The device evaluation found the emergency light error panel was flashing- the incident occurred due to the lamp burnout of the emergency light.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus that on the visera elite xenon light source, the emergency light error panel was flashing.The reported problem was found during inspection for use, and there was no delay.The facility finished the therapeutic procedure with a similar device.There were no reports of patient harm.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 4 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a definitive root cause of the lamp burnout related to the emergency light could not be identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA ELITE XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17294344
MDR Text Key318760502
Report Number3002808148-2023-06902
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170301995
UDI-Public04953170301995
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K111425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 08/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-S190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2023
Was the Report Sent to FDA? No
Date Manufacturer Received07/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/11/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-