MAUDE Adverse Event Report: AESCULAP AG LIGATURE CLIP 12 MAG.= 144 PCS.; HANDHELD PRODUCTS & LIGATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG LIGATURE CLIP 12 MAG.= 144 PCS.; HANDHELD PRODUCTS & LIGATION Back to Search Results
Model Number PL572T
Device Problem Material Fragmentation (1261)
Patient Problem Foreign Body In Patient (2687)
Event Date 06/01/2023
Event Type  malfunction  
Event Description
It was reported that there was an issue with the product pl572t - ligature clip 12 mag.= 144 pcs.According to the complaint description, the plastic end of a titanium clip bar broke off during use and was not found.This event occured during a surgery for kidney donation.A fragment remained in situ.Additional information was not provided nor available.Additional patient information is not available.The malfunction is filed under aag reference (b)(4).
 
Manufacturer Narrative
Investigation results: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 3(5) x probability 2(5) of occurrence) according to din en iso 14971 is still acceptable.Explanation and rationale: there is the possibility that the tip of the titanium clip cassette became pre-damaged during the assembly process.According to the applicable instruction for use, the user shall make sure that the tip is not damaged during that process.However, due to a lack of data and without the product we cannot determine an exact conclusion and root cause for the reported issue.According to the quality standard, a production error and a material defect can most probably be excluded.If further investigations are required, the product should be provided for examination.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.Based upon the investigations results a capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGATURE CLIP 12 MAG.= 144 PCS.
Type of Device
HANDHELD PRODUCTS & LIGATION
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key17299614
MDR Text Key319137448
Report Number9610612-2023-00167
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K081031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPL572T
Device Catalogue NumberPL572T
Device Lot Number52794834
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/19/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/09/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-