Model Number 2001-65-000 |
Device Problems
Crack (1135); Material Deformation (2976); Naturally Worn (2988); Scratched Material (3020)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/30/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that sales rep has 6 femoral head impactor white tips that are damaged and need replaced.All 6 femoral head impactors were cracked or bent on the white tip from forceful impaction causing wear over time.There was no surgical delay or patient consequences.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint (b)(4).Investigation summary: an analysis of the product could not be performed, since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed, as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information, monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product or product id numbers are received at a later date, the investigation will be updated as applicable.Added: b5, h6: (medical device problem code).
|
|
Event Description
|
Additional information was received.And stated, that the product's etch was, due to usage over time.
|
|
Search Alerts/Recalls
|
|