Model Number 8002950
Device Problems Failure to Pump (1502); Pumping Problem (3016); Component Misassembled (4004)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the during support, a faulty pcb assembly and a re-circulating water problem were found.No patient was involved.
 
Manufacturer Narrative
B3: date of event is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
Other text: d10.Device available for evaluation; h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.One device was received.Visual inspection noted the return tube was cracked and the printed circuit board (pcb) had evidence of fluid ingression.Functional testing confirmed the complaint as the device was not entering "operation mode" (green led on display is not lit) it powered on but would not run.The root cause was due to a cracked return tube which allowed fluid to leak onto pcb causing it to fail from a design issue.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced the return tube and pcb to correct the reported problem.Replaced the o-rings per preventative maintenance (pm).Device passed all functional and delivery tests.
 
Manufacturer Narrative
Other, other text: on review it was found that the information in this report is a duplicate of 3012307300-2023-07169.Please disregard 3012307300-2023-07149 and all files associated with it as all information has been submitted on 3012307300-2023-07169.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 FAST FLOW FLUID WARMER
Type of Device
WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17309580
MDR Text Key319394422
Report Number3012307300-2023-07149
Device Sequence Number1
Product Code BSB
UDI-Device Identifier50695085829506
UDI-Public50695085829506
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK020043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 01/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8002950
Device Catalogue Number8002950
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/23/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-