• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM MEXICO ADVANTA 2 FRAME; BED, MANUAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM MEXICO ADVANTA 2 FRAME; BED, MANUAL Back to Search Results
Model Number P1190A000248
Device Problem Activation Problem (4042)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/16/2023
Event Type  malfunction  
Manufacturer Narrative
The hillrom technician found the cpr cable needed to be replaced.Per the hillrom service manual the advanta¿ 2 should be subject to an effective maintenance program.This will help make sure of a long, operative life for the advanta¿ 2 bed.The preventative maintenance will help to reduce downtime due to excessive wear failures.An annual service of the bed is advised in order to maintain its characteristics and performance.Examine the handles, cables, and cpr mechanism on the head motor.Make sure the screws are installed and fully tightened.Raise the head section to the high position, then activate one of the cpr releases.Make sure the head section lowers.Adjust the cpr cable as necessary.Do the same test on the other side of the bed.Release the cpr handles and make sure the mechanism locks correctly by operating the head up control for a few seconds.The head section should rise.Replace the head section motor as necessary.A search of the hillrom maintenance records showed hillrom performed preventative maintenance on this bed on (b)(6) 2023.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the cpr cable to resolve the reported event.Based on this information, no further action is required.
 
Event Description
Hillrom received a report from a hillrom technician stating the cpr was inoperative.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTA 2 FRAME
Type of Device
BED, MANUAL
Manufacturer (Section D)
HILL-ROM MEXICO
ave.del telefono no. 200 col. huinala
apodaca, nuevo leon 66640
MX  66640
Manufacturer Contact
estephania cardenas
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key17320233
MDR Text Key319678079
Report Number3006697241-2023-00065
Device Sequence Number1
Product Code FNJ
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 07/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP1190A000248
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/21/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-