Catalog Number 110003451
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/15/2023
Event Type  Injury  
Event Description
It was reported that during the procedure, the cup got stuck to the impactor.There was no harm or health consequences to the patient.It was reported that no further information is available.
 
Manufacturer Narrative
(b)(4) multiple mdr reports were filed for this event, please see associated reports: 0001825034-2023-01619.D10: cat #: 110010243 / g7 osseoti 3 hole shell 50mm d / lot #: 65698272 product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2023-01824.Visual examination of the returned product identified the devices are stuck together and could not be disassembled.There are indentations to the strike plate of the inserter.The threaded shaft hex feature has been stripped.This complaint was confirmed based on the returned devices.Dhr was reviewed and no discrepancies related to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 STR MODULAR SHELL INSERTER
Type of Device
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17325102
MDR Text Key319482579
Report Number0001825034-2023-01620
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/14/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number110003451
Device Lot Number970130
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/27/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/01/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE; UNK THREADED SHAFT
Patient SexFemale
-
-