Model Number N/A |
Device Problem
Packaging Problem (3007)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/20/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4) the product has been received by zimmer biomet.The investigation is currently in process.Once the investigation has been completed, a follow-up report will be submitted.
|
|
Event Description
|
It was reported that the sterile packaging was damaged.No patient involvement was reported.There is no additional information available at the time of this report.
|
|
Event Description
|
There is no update to the prior event description provided.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4 g3 g6 h2 h3 h6 h10 visual evaluation of the returned product found the sterile blister is cracked.Sterility has been breached.This complaint has been confirmed by evaluation of the returned product.Review of the device history records identified no deviations or anomalies during manufacturing.Medical records were not provided.The condition of the device when it left zimmer biomet is considered conforming to specification.The root cause of the reported event can be attributed to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|