Model Number N/A |
Device Problems
Material Frayed (1262); Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/16/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
An investigation into the reported event has been initiated under (b)(4).Once the investigation has been completed, a follow up report will be submitted with the investigation results and any actions taken by the manufacturer.
|
|
Event Description
|
It was reported that after surgery, the plug that goes into power unit came apart when unplugging it.There was no patient involvement.The power cord or wires were not frayed but the plug end had come apart.The metal wires were exposed.Due diligence is complete, no further information is available.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the rpms were not checked due to a damaged plug harness cord.The plug harness was replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
There is no additional information available.
|
|