Device evaluated by mfr.: the returned product consisted of a watchman delivery system with the implant in the sheath and blood in the device.The implant, core wire, tip, sheath, and hub/valve were microscopically and visually examined.Inspection revealed numerous kinks in the sheath.There was tip damage as there were abrasions on the inside of the sheath and tip, the tri-cuts were flared, and the distal marker band was kinked.There was implant anchor damage.The anchor and tip damage is consistent with deploying and recapturing the implant.Functional testing was completed by attaching a syringe filled with water while positive and negative pressure was applied with the valve open and closed.The device was not able to be flushed properly.The device was not able to backflush, and air was not able to purged from the device.The valve was removed and microscopically examined.The silicone valve was found to be damaged.Product analysis confirmed the reported event of difficulty flushing as the device was difficult to flush and damage to the valve was observed during analysis, which could have contributed to difficulty with the flushing of the device.Product analysis confirmed the reported event of marker band damage as the distal marker band was kinked.
|