ST PAUL CADD CASSETTE RESERVOIRS - FLOW STOP; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7302-24
Device Problems Leak/Splash (1354); Gas/Air Leak (2946)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported the disposable had two leaks occurred.Second pump leak came in this week.Dislodged from the pigtail and extension set.Equashield ll-2 lot 2212817a was possibly used during this infusion.No patient injury.No additional information available.Device is returning for investigation.There were no allegations against the pumps but they will be sent in for investigation.Devices are all covered in chemo.
 
Manufacturer Narrative
One sample was returned for investigation.Visual inspection found no damages or defects to the device.Functional testing of the device occurred.No leak was detected during testing.The reported complaint is not confirmed.No root cause determined.No actions taken.A review of the device history records shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD CASSETTE RESERVOIRS - FLOW STOP
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17333392
MDR Text Key319225561
Report Number3012307300-2023-07219
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10610586027239
UDI-Public10610586027239
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K040636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/17/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21-7302-24
Device Catalogue Number21-7302-24
Device Lot Number4354595
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Date Manufacturer Received09/30/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-