EDWARDS LIFESCIENCES INSPIRIS RESILIA AORTIC VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number 11500AJ19
Device Problem Fluid/Blood Leak (1250)
Patient Problem Insufficient Information (4580)
Event Type  Injury  
Event Description
It was reported that a patient with a 19mm 11500aj aortic valve underwent a valve-in-valve procedure after an implant duration of three (3) years, eleven (11) months due to aortic regurgitation.The procedure was performed with a spien3 ultra transcatheter valve.The patient status was reported as "under treatment".The device was not returned for evaluation as it remained implanted.
 
Manufacturer Narrative
H10: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Manufacturer Narrative
The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.Regurgitation which develops progressively over time can be due to a number of issues including patient related factors or structural valve deterioration.Structural valve deterioration (svd) can, and typically does, lead to chronic central leaks over a period of time.Svd is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, noncalcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes may occur singularly or concomitantly.Tissue degeneration-related structural deterioration, either calcific or non-calcific, are common chronic failure modes for this type of bioprosthetic heart valve.Operational mechanical stress and biological factors are generally believed to be the major contributors to the non-calcific bioprosthetic tissue degeneration.The most likely cause is patient factors, including a history of dialysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRIS RESILIA AORTIC VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
irvine, CA 92614
9492506615
MDR Report Key17337260
MDR Text Key319182391
Report Number2015691-2023-14639
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/17/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date03/04/2021
Device Model Number11500AJ19
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/16/2023
Date Device Manufactured03/05/2019
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Life Threatening; Hospitalization;
Patient Age86 YR
Patient SexFemale
Patient Weight54 KG
-
-