Date of event is estimated.Additional components potentially involved in the event include: common device name: drg lead, model: mn10450-50a, udi: (b)(4), serial: (b)(6), batch: 8713227.Common device name: drg lead, model: mn10450-50a, udi: (b)(4), serial: (b)(6), batch: 8713227.Common device name: drg lead, model: mn10450-50a, udi: (b)(4), serial: (b)(6), batch: 8713227.
|
It was reported that, following an implant procedure, the patient was hospitalized due to multiple issues, including seizures and not being able to breathe on their own.A manufacturer representative met with the patient and it was determined the patient was hospitalized due to low blood sugars.Patient's therapy is currently on and providing coverage.
|
Based on the information provided a device problem was not identified, as a result a conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.
|