Model Number 912082
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/11/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).G2: foreign: (b)(4).Visual examination of the returned product identified one inserter prong has fractured.The device was returned with no sutures or anchors.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the instrument anchors pulled out during the procedure.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JGRKNT 1.0MM MINI 3-0 NDLS
Type of Device
FASTNER, FIXATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17341428
MDR Text Key319418963
Report Number0001825034-2023-01618
Device Sequence Number1
Product Code DZL
UDI-Device Identifier00880304523418
UDI-Public(01)00880304523418(17)271012(10)0002475765
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K110879
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number912082
Device Catalogue Number912082
Device Lot Number0002475765
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/12/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-