SYNTHES GMBH BATTERY RECIPROCATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Model Number 530.715 |
Device Problems
Complete Loss of Power (4015); Intermittent Energy Output (4025)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that during service and evaluation, it was determined that the battery reciprocator device had intermittent operation and failed pre-test for check function of the device and check the oscillation frequency with the frequency meter.It was noted in the service order that the device was broken and had no power.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.All available information has been disclosed.If additional information should become available, an additional report will be submitted accordingly.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.During evaluation it was determined that the initial reported condition was not confirmed.Therefore, the assignable root cause was not determined.However, the malfunctions found during service and evaluation have been confirmed.The assignable root cause was determined to be due to user error.Udi:(b)(4).
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Upon further evaluation of the returned device, it was found that the device also had corrosion/rusting/pitting and the moving parts did not move smoothly.H6: the assignable root cause was reported as user error in the initial report and has been updated to improper maintenance.
|
|
Search Alerts/Recalls
|
|
|