Catalog Number 8606500 |
Device Problems
Gas Output Problem (1266); Failure to Deliver (2338); Protective Measures Problem (3015)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/23/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during use of automatic ventilation, when connecting the inoflo connector to the inspiratory side, the device alarmed for "ventilator fail" and switched to manual ventilation.Switching back to automatic ventilation was, as further reported, not possible.There was no patient consequences reported.
|
|
Manufacturer Narrative
|
The investigation has just started; results will be provided in a follow-up report.H3 other text : on-going.
|
|
Manufacturer Narrative
|
The electronic device logfile was provided for investigation.Unfortunately, the user section of the logfile was already overwritten for the given date (2023-06-23).Thus, the case in question could not be reconstructed in detail.According to the device log records, it could be confirmed that on the date of event, the ventilator performed an emergency shutdown due to a detected overpressure at the patient end of the circuit leading to a pressure peak.Most likely, the overpressure situation was caused by the surgical therapy, but also a patient coughing against the ventilator could explain the pressure peak.A connection between the usage of the inoflo system in combination with the apollo and the reported symptom could neither be confirmed nor excluded on the basis of the available information.To prevent from damages, the system is designed to shut down automatic ventilation and to alert the user to this condition by means of a corresponding alarm.Manual ventilation and the monitoring functions remain available to the full extent.Dräger finally concludes that the device responded as specified upon the detected situation with an autonomous shutdown while changing mode to man/spont (safety mode) accompanied by an audible and visible "ventilator fail" alarm.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|
|
Event Description
|
It was reported that during use of automatic ventilation, when connecting the inoflo connector to the inspiratory side, the device alarmed for "ventilator fail" and switched to manual ventilation.Switching back to automatic ventilation was, as further reported, not possible.There was no patient consequences reported.
|
|
Search Alerts/Recalls
|
|