Catalog Number 261221
Device Problem Fracture (1260)
Patient Problem Insufficient Information (4580)
Event Date 07/04/2023
Event Type  malfunction  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A physician reported the portion of the sleeve of a perforator (id 261221) cracked during surgery.It occurred when the third burr hole was drilled.It is unknown if the part fall into the surgical site, however, the cracked part was completely removed.According to information provided, it is unknown the drill used with the perforator.It is also unknown if the drill is electric or pneumatic and if the perforator clicks in place in the drill and if the recommended spring tests were performed between each burr hole.The perforator was used for brain tumorectomy, no significant surgical delay.
 
Event Description
N/a.
 
Manufacturer Narrative
The codman perforator (261221) was returned for evaluation.Device history record (dhr) - there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Failure analysis: the perforator unit was inspected using the unaided eye: unit was received disassembled and soiled with organic material.The "ifu" testing procedure was performed with no observed anomalies after sleeving and reassembling the unit as it could not be tested in the received condition.No new parts were needed, unit was reassembled with the parts sent back.Functional testing was performed using the same protocol it underwent at finished goods testing prior to release and the unit was found to perform as intended and fulfilled the acceptance criteria.The complaint could not be verified through failure analysis.Root cause: the root cause is undetermined and was unable to be confirmed in the complaint evaluation.The potential causes of failure include user misuse.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
RAYNHAM
325 paramount drive
325 paramount drive
raynham MA
Manufacturer (Section G)
RAYNHAM
325 paramount drive
raynham MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key17350245
MDR Text Key319709055
Report Number3014334038-2023-00112
Device Sequence Number1
Product Code HBF
UDI-Device Identifier10381780513599
UDI-Public10381780513599
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/19/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number261221
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/17/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-