Catalog Number 51-104170
Device Problems Manufacturing, Packaging or Shipping Problem (2975); Packaging Problem (3007)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/27/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2023-01673.0001825034-2023-01674.0001825034-2023-01676.G2: japan.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during an inspection of the items currently stocked, the sterile package was found to be damaged.There was no patient involvement.It was reported that no further information is available.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4; b5; g3; h2; h3; h6.Visual evaluation of the returned product identified damage to the sterile packaging (blister).Sterility has not been breached.The reported event has been confirmed by evaluation of the returned product.Dhr was reviewed and no discrepancies related to the reported event were found.The condition of the device when it left zimmer biomet is considered conforming to specification.The root cause of the reported event can be attributed to transit damage and a packaging design issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 T1 PPS HO 17X154MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17350919
MDR Text Key319308767
Report Number0001825034-2023-01675
Device Sequence Number1
Product Code LZO
UDI-Device Identifier00880304499256
UDI-Public(01)00880304499256(17)300923(10)6858280
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K200196
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/19/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number51-104170
Device Lot Number6858280
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2023
Was the Report Sent to FDA? No
Date Manufacturer Received10/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/23/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-