• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA III DUODENOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA III DUODENOVIDEOSCOPE Back to Search Results
Model Number TJF-Q190V
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned as part of the asset return process.In addition to the dirt found inside the light guide lens, other evaluation findings are as follows: the scope body and the light guide lens had a crack; water tightness was lost due to damage on the upward/downward knob, the right/left knob and a pinhole on the connecting tube; there were scratches on the grip, the switch box, and the electrical contract of the endoscope connector; the air/water cylinder and the suction cylinder were discolored; the label on the suction connector was damaged; the objective lens had a chip; the adhesive around the objective lens was worn; there was corrosion around the forceps elevator; and the coating on the universal cord was peeling.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
 
Event Description
The duodenovideoscope was returned as part of the asset return process.During routine inspection of the device by olympus, dirt was found inside the light guide lens.There was no procedural or patient involvement associated with this event.This mdr is being submitted to capture the reportable malfunction found during the device evaluation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported event is unable to be determined.However, the cause of the event is likely due to physical stress caused by hitting distal end of the device or dropping distal end, or chemical stress caused by chemical solution used, lg-lens glue peeled off and moisture entered inside of lg-lens and corroded.The event can be detected by following the instructions for use (ifu) section: operation manual 3.3 inspection of the endoscope shows how to detect the event.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17360201
MDR Text Key319419132
Report Number9610595-2023-10343
Device Sequence Number1
Product Code FDT
UDI-Device Identifier04953170452024
UDI-Public04953170452024
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K220587
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTJF-Q190V
Device Catalogue NumberN5992941
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/08/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-