ST PAUL CADD CASSETTE RESERVOIRS - FLOW STOP; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7301-24
Device Problem Gas/Air Leak (2946)
Patient Problem Insufficient Information (4580)
Event Date 05/31/2023
Event Type  malfunction  
Manufacturer Narrative
Other, other text: a supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported the disposable exhibits "a risk of leakage" at the disposable tubing near the clamp.The patient noticed a bulge bordering on rupture, and when she engaged the disposable, the device emitted a beep-type alarm with no message.Device is available for return.No credit or replacement requested.The patient did not keep the device but a photo is available.
 
Manufacturer Narrative
No product was returned.The reported issue cannot be confirmed as no product was returned for investigation.If the product is returned, this complaint will be reopened for further investigation.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.G2 email is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD CASSETTE RESERVOIRS - FLOW STOP
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17363076
MDR Text Key319520987
Report Number3012307300-2023-07323
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10610586027215
UDI-Public10610586027215
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K040636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number21-7301-24
Device Catalogue Number21-7301-24
Device Lot Number4334116
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/19/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age42 YR
Patient SexFemale
-
-