Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h10.The mayfield modified skull clamp (a1059) was returned for evaluation: device history record (dhr) - the dhr was reviewed, and no anomalies related to the reported failure was observed.Failure analysis - the investigation of the returned device found no device deficiencies that would have contributed to the reported complaint.No maintenance is currently required.Unrelated to the complaint issue, the swivel lock was found to have movement in the assembly and a residue buildup was present.Additionally, the unit required cleaning and replacement of worn parts from routine use and wear.Some worn off internal were replaced, general maintenance and cleaning were performed and the unit has been subjected to a successful functional check.Root cause - evaluation found no device deficiencies that would have contributed to the reported complaint.The probable root cause of the reported complaint is improper or suboptimal placement of the skull clamp on the patient.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|