Other, other text: for all enquiries or follow-up questions related to the record, do (b)(6)additional information: a1, b3, b5 and h6 - health effects codes.H3 and h6 - evaluation codes: updated.Device evaluation: one device was returned for investigation.Visual inspection found the device was leaking from drain valve.Device had an old style float switch intact.Functional testing found the plunger gets stuck during the pin gauge test on the air detector.Unable to duplicate ramp time as the temp performed during cold start: in 60 seconds temp reached 30 degrees celsius.And by 3 mins temp reached 41.7 degrees celsius.Display temp and tempcheck temp were stable for 45 mins, but both out of spec at 41.9 degrees celsius; exceeding the temp.The complaint was confirmed.Root cause was attributed to assembly error.The optical cable is in the way of the solenoid to go up and down.Needs calibration to the unit.Service history review identified there was no indication that the complaint was related to a service of the device within the review period.Re-assembled to re-tight and re-installed the optical cable.Performed calibration to the unit.Rreplaced the o-rings and the fan guard for pm.Replaced the drain valve and float switch.
|