Model Number 8002950
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/28/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the pressure chambers orange dial for the pressure gauge was tucked in behind the meter.This is causing the pressure reading to not be visible to the users.No patient involvement.
 
Manufacturer Narrative
G2 email is: (b)(4).No product was returned.Reviewed manufacturing bill of materials (bom) of the trauma tower and pressure chamber.Pressure chambers and trauma tower are built, packaged, and shipped separately.The manufacturing bom of tower does not include pressure chamber as part of the packaging process.The failed pressure chamber was returned on complaint (b)(4) where the investigation determined incorrect assembly resulted in the attachment tab of the pressure gauge to break off and dislodge.No fault found with the tower.Pressure chamber are built and shipped separately.A replacement pressure chamber request was processed to the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 FAST FLOW FLUID WARMER
Type of Device
WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17368439
MDR Text Key319759952
Report Number3012307300-2023-07476
Device Sequence Number1
Product Code BSB
UDI-Device Identifier50695085829506
UDI-Public50695085829506
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
BK020043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8002950
Device Catalogue Number8002950
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/31/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/08/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-