• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CRITICAL CARE AB SUPPORT ARM 177; SUPPORT, ARM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CRITICAL CARE AB SUPPORT ARM 177; SUPPORT, ARM Back to Search Results
Model Number SUPPORT ARM 177
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/12/2023
Event Type  malfunction  
Event Description
It was reported that the ventilator's support arm was broken.There was no patient harm reported.Manufacturer's ref.#:(b)(6).
 
Manufacturer Narrative
According to our field service engineer, the customer replaced the support arm and purchased it from another supplier.No further information was provided and no further issues have been reported after the event.In the installation manual for the support arm 176/177 in the specification of the maximum capacity, it is stated how many kilograms can be loaded depending on the angle of use of the accessories (max.Approx.3 kg).If the support arm will be broken, it may lead to stop of ventilation (extubation) or injury.The root cause cannot be established by this investigation.
 
Event Description
Manufacturers ref: # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPPORT ARM 177
Type of Device
SUPPORT, ARM
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
caroline kabbabe
roentgenvagen 2
solna 
MDR Report Key17370952
MDR Text Key319506705
Report Number8010042-2023-01420
Device Sequence Number1
Product Code IOY
Combination Product (y/n)N
Reporter Country CodeIT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSUPPORT ARM 177
Device Catalogue Number6481720
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received11/06/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-