• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APOLLO ENDOSURGERY, INC OVERSTITCH¿ ENDOSCOPIC SUTURING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APOLLO ENDOSURGERY, INC OVERSTITCH¿ ENDOSCOPIC SUTURING SYSTEM Back to Search Results
Model Number ESS-G02-160
Device Problem Material Perforation (2205)
Patient Problem Laceration(s) of Esophagus (2398)
Event Date 02/08/2023
Event Type  malfunction  
Event Description
The complaint was reported to apollo from fda under # 0700220000-2023-8103.The suture cinch malfunctioned and would not cut and release from the end of the suturing in the esophagus.The device had to be dismantled to be removed from the patient esophagus.The patient is doing well.
 
Manufacturer Narrative
Combined medwatch submitted to the fda on 21/jul/2023.A review of the device labeling notes the following: the current overstitch¿ endoscopic suturing system (ess) instructions for use (ifu) addressed the known and potential event of "stomach perforation/erosion" as follows: the apollo endosurgery overstitch¿ endoscopic suture system (ess) is intended for endoscopic placement of suture(s) and approximation of soft tissue.Contraindications include those specific to use of an endoscopic suturing system, and any endoscopic procedure, which may include, but not limited to, the following: this system is not for use where endoscopic techniques are contraindicated.This system is not for use with malignant tissue.Adverse events possible complications that may result from using the endoscopic suturing system include, but may not be limited to: pharyngitis / sore throat, nausea and / or vomiting, abdominal pain and / or bloating, hemorrhage, hematoma, conversion to laparoscopic or open procedure stricture, infection / sepsis, pharyngeal, colonic and/or esophageal perforation, esophageal, colonic and/or pharyngeal laceration, intra-abdominal (hollow or solid) visceral injury aspiration, wound dehiscence, acute inflammatory tissue reaction death.Note: any serious incident that has occurred in relation to the device should be reported to apollo endosurgery (see contact information at the end of this document) and any appropriate government entity.Additional information: the device has not been returned for analysis, and after multiple attempts to gather more information from the reporter, no additional information has been received.The investigator determined a device history record (dhr) review is not possible for this complaint, as attempts at gathering the device serial/lot number were unsuccessful.Device evaluation summary: assessment of the device involved in this complaint was not possible, and it has not been possible to determine the root cause for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVERSTITCH¿ ENDOSCOPIC SUTURING SYSTEM
Type of Device
ENDOSCOPIC SUTURING SYSTEM
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC
1120 s. captail of texas hwy
bldg 1, ste 300
austin 78746
Manufacturer (Section G)
APOLLO ENDOSURGERY COSTA RICA, SRL
coyol free zone
building b 13.3
alajuela, cs CRI
CS   CRI
Manufacturer Contact
adriana russell
1120 s. captail of texas hwy
bldg 1, ste 300
austin 78746
5122795114
MDR Report Key17371081
MDR Text Key319516026
Report Number3006722112-2023-00145
Device Sequence Number1
Product Code OCW
UDI-Device Identifier10811955020664
UDI-Public(01)10811955020664
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081853
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberESS-G02-160
Device Catalogue NumberESS-G02-160
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/29/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-