Medtronic received information regarding a navigation device being used during a sacroiliac and thoracolumbar procedure.It was reported that the site had experienced an alleged inaccuracy.While in surgery the site took a spin and was verifying, but reported an inaccuracy that was 1 cm deep.They took 2 spins with the same result.The site moved to fluoro to continue the surgery.When the medtronic representative (rep) was testing the system, he observed similar inaccuracies.After spinning when verifying the midline he found it was about 1-2 mm deep, but became 4-5 mm deep when moving lateral on either side.He took a second spin while making sure that nothing moved and was still seeing about 2-3 mm deep.He then tried a point merge registration off of the second spin and got an error metric of 1.0, but still felt it was 1 mm deep on the midline and 2-3 mm lateral.There was no delay or patient impact report ed.
|
Section d references the main component of the system.Other medical products in use during the event include: bi71000218, lot number: unknown h3/h6: the system was serviced and there were no failures found.Codes b01, c19, and d14 apply.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|